Dual Energy CT huvud Lars Marklund Siemens Healthcare
SOMATOM Definition Flash Dual Source Dual Energy for Neuro Singel Source CT Huvud Akademiska 2014-08 Scanprotokoll: 120 kv, Q ref: 380mAs CTDI: 50 mgy, DLP: 800 mgy*cm Kollimering: 128 x 0,6 mm Pitch: 0,55 Rotationstid: 1 s Undersökningstid: 10 s Rekonstruktioner: J37s 4mm ax, cor, sag SAFIRE 3 Dual Energy CT Huvud Akademiska 2014-08 Scanprotokoll: A: 80kV, Q ref: 700 mas, B: Sn140kV 350 mas CTDI: 45 mgy, DLP: 700 mgy*cm Kollimering: 2 x 40 x 0,6 mm Pitch: 0,55 Rotationstid: 1 s Undersökningstid: 20 s Rekonstruktioner: J37s 4mm ax, cor, sag SAFIRE: 3 Fast DE: 0,7
SOMATOM Definition Flash Dual Source Dual Energy for Neuro CT Huvud 120 kv Dual Energy CT Huvud
SOMATOM Definition Flash Dual Source Dual Energy for Neuro Dual Energy CT Head with contrast kontrast Scanparameters: Same as DE Head protocol without iv contrast. Reconparametrar: 4 mm MPR (Pacs) Kernel: J37s, Safire 3 1 mm MPR (Sy.via) Kernel: D34s Viktning av mixad MPR serie: FAST DE 0,5-0,8 Grad av iteration: SAFIRE 1-3 Dual Energy applikationer I syngo.via Brain Hemorrhage Utsöndras kontrastmedel i en gammal blödning? Virtuellt nativ serie mg jod / ml vävnad Metallartefaktsreduktion (Främst för skelett) Boneremoval Optimum Contrast
Brain Hemorrhage Copyright Siemens AG 2012. All rights reserved. Internal only. Restricted Siemens AG 2013. All rights reserved. siemens.com/answers Answers for life.
Brain Hemorrhage Copyright Siemens AG 2012. All rights reserved. Internal only. Restricted Siemens AG 2013. All rights reserved. siemens.com/answers Answers for life.
120kV Head (true) non contrast nc DE Virtual Head
Copyright Siemens AG 2012. All rights reserved. Internal only. Restricted Siemens AG 2013. All rights reserved. siemens.com/answers Answers for life.
CARE Dose4D Stråldosen beräknas med tre parametrar
CARE Dose4D Stråldosen beräknas med tre parametrar
Noise Variations (pitch artifact) In abdomen scans it may happen that the noise e.g. in the liver changes spiralshaped with the z position. This is particularly visible in MPRs where increased noise is imaged as increased brightness: z
Noise Variations In abdomen scans it may happen that the noise e.g. in the liver changes spiralshaped with the z position. This is particularly visible in MPRs where increased noise is imaged as increased brightness: z
Noise Variations The origin of this effect is the fact that in spiral mode the dose is not equally distributed. With a 3D backprojection, this dose distribution will be mapped into a noise distribution in the volume.
Noise Variations Pitch 1.0: in a coronal plane we have regions with multiplicity 1, 2 or 3. 3x 1x z 1x 3x 3x 1x
Noise Variations Pitch 0.6: in a coronal plane we have mostly regions with multiplicity 2 How to prevent the artifact: The artifact is only visible if low contrast regions (e.g. liver) are read with a narrow (soft-tissue) window. It will not be noticeable e.g. in a CTA. In exams where the effect is relevant, the pitch should be lowered, e.g. to 0.6. z
Left: 100 kv, CTDIvol: 5,3 mgy (Care kv) Right: 120 kv, CTDIvol: 9,45 mgy